Calibration Kits
Reference kits for NIST, NRC, US Pharmacopoeia and European Pharmacopoeia spectrophotometer validation. All standards are NIST traceable and prepared using NIST guidelines. Custom kits are available upon request.
FNIR Photometric Accuracy & Linearity Kit (10%T, 20%T, 30%T) (700-3000nm)
FNIR Photometric Accuracy & Linearity Kit (10%T, 20%T, 30%T) (700-3000nm)
Kit of three filters for validating photometric accuracy and linearity in the range of 700-3000nm
This kits includes the following filters.
Note: Performance of calibration standards of this kit entirely complies with performance requirements of SRM NIST 2031a kit, while absorbance/transmittance values of reference standards of this kit are identical to some reference standards of NIST 2031a.
Capabilities:
- Photometric accuracy in the UV and VIS range (700-3000nm)
Product Specs:
- Optical densities: 1.0au (10%), 0.7au (20%), 0.5au (30%)
- Wavelength Range: 700-3000nm. Certified wavelength points include 700, 1100, 1700, 2700 and 3000nm. Firefly can provide alternative points in the 700-3000nm range upon request or full scan data for informational purposes.
- Tolerances: +/-0.0022 au
- External Dimensions: 12.5 x 12.5 x 45 mm
Included with Kit:
- Instruction manual
- Certificates of calibration with NIST traceability
- Photometric data tables are available for qualified customers. Please contact us for more information.