Calibration Kits
Reference kits for NIST, NRC, US Pharmacopoeia and European Pharmacopoeia spectrophotometer validation. All standards are NIST traceable and prepared using NIST guidelines. Custom kits are available upon request.
UV/VIS Photometric & UV/VIS Wavelength Accuracy Kit
UV/VIS Photometric & UV/VIS Wavelength Accuracy Kit
Standard calibration kit for checking photometric accuracy in the range of 200-700nm and wavelength accuracy in the range of 241-637nm.
This kit includes the following filters:
- FUV-0.3 (50%T)
- FUV-0.5 (30%T)
- FUV-1.0 (10%T)
- WAV-1 UV/VIS Solid State Holmium Oxide Filter
- A30 Blank holder
- Case G6
Capabilities:
- Photometric accuracy in the VIS range (200-700nm) - Data points of 250, 280, 340, 360, 400, 465, 500, 546.1, 590 and 635nm included on certificate.
- Wavelength accuracy in the UV and VIS range (241.5-637.5nm) - Data points of 241.5, 279.3, 287.6, 333.8, 360.8, 385.8, 418.5, 453.4, 459.9, 536.4 and 637.5nm included on the certificate.
Product Specs:
- Tolerance for photometric accuracy filters: +/- 0.0034 au
- Tolerance for wavelength accuracy filters: +/- 0.2nm
- External dimensions of individual filters: 12.5 x 12.5 x 45.0 mm
Benefits of Firefly's Solid-State Calibration Standards:
- Highly accurate and consistent spectrophotometer calibration every time.
- Never has to be replaced.
- Never has to be recalibrated if handled and stored properly.
- Scratch-resistant coating protects optical surfaces.
- Can be used in any spectrophotometer (with standard 12.5 x 12.5 x 45.0 mm holder)
- Surpasses NIST requirements and tolerances.
- NIST-traceable.